Temperature and Humidity
Source: https://emi.qcc.gov.ae/EMI-Laboratories/Temperature-and-Humidity-Lab
Archived: 2026-04-23 17:32
Temperature and Humidity
Temperature
The Temperature Laboratory realises the International Temperature Scale of 1990 (ITS-90) through a set of fixed point cells (TPW, Hg, Ga, In, Sn, Zn, Al, Ag and Cu) and a set of standard platinum resistance thermometers (SPRTs).
Calibration services provided are:
Standard Platinum Resistance Thermometers
Calibration of Standard Platinum Resistance Thermometers directly at the fixed points of TPW, Hg, Ga, In, Sn, Zn and Al, covering the temperature range from -38.8344 °C to 660.323 °C.
Calibration Service
Range
Calibration and Measurement Capability (CMC)
SPRT Calibration at Fixed Points
-38.8344 °C
0.6 mK
0.01 °C
0.3 mK
29.7646 °C
0.6 mK
156.5985 °C
1.7 mK
231.928 °C
1.6 mK
419.527 °C
1.9 mK
660.323 °C
16 mK
SPRT Calibration at Fixed Points Sub Range
-38.8344 °C to 0.01 °C
0.7mK
0.01 °C to 29.7646 °C
0.7 mK
29.7646 °C to 156.5985 °C
1.8 mK
156.5985 °C to 231.928 °C
2 mK
231.928 °C to 419.527 °C
3 mK
419.527 °C to 660.323 °C
3mK to 16 mK
Platinum Resistance Thermometers
Calibration of Platinum Resistance Thermometers in comparison with SPRTs, using liquid baths to create uniform and stable temperature conditions, covering the range from -80 °C to 250 °C.
Calibration Service
Range
CMC
Platinum Resistance Thermometers
-80 °C to 80 °C
0.01
> 80 °C to 150 °C
0.015
> 150 °C to 250 °C
0.02
419.527 °C
0.005
Temperature Sensors with Display Unit/
Temperature sensors with display units are calibrated in liquid baths and in block calibrators in comparison with reference thermometers.
Calibration Service
Range
CMC
Resistance Thermometers With Display Unit
-80 °C to 80 °C
0.015 °C
> 80 °C to 150 °C
0.020 °C
> 150 °C to 250 °C
0.025 °C
> 250 °C to 400 °C
0.6 °C to 0.8 °C
Thermocouples With Display Unit
-80 °C to 150 °C
0.2 °C
> 150 °C to 250 °C
0.4 °C
> 250 °C to 400 °C
0.6 °C to 0.8 °C
Liquid In Glass Thermometers
Liquid in glass thermometers are calibrated in liquid baths from -60 °C to 250 °C by comparison with reference thermometers.
Calibration Service
Range
CMC
Liquid In Glass Thermometers
-60 °C to 250 °C
0.02 °C to 0.05 °C
Air Temperature Sensors
Calibration of air temperature sensors and data loggers in the range of 10°C to 70°C, in an air chamber.
Calibration Service
Range
CMC
Air Temperature Sensors
10 °C to 70 °C
0.10 °C to 0.30 °C
Infrared thermometers
IR Thermometers are calibrated by comparison using black body sources in the range from -150 °C to 500 °C.
Calibration Service
Range
CMC
Infrared thermometers
-15 °C to 500 °C
0.6 °C to 2.6 °C
Temperature Block Calibrators
Temperature block calibrators are calibrated using platinum resistance thermometers in the range from -30 °C to 400 °C.
Calibration Service
Range
CMC
Temperature Block Calibrators
-30 °C to 400 °C
0.2 °C
Liquid Baths
Calibration of liquid baths is carried out from -80 °C and 250 °C. The temperature profile of the bath can also be measured.
Calibration Service
Range
CMC
Liquid Baths
-80 °C to 250 °C
0.02 °C to 0.03
°C
Climatic Chambers
Climatic chambers are calibrated in the range from -40 °C to 140 °C. The temperature profile of the chamber can also be measured using 9 PRT sensors distributed in the working volume of the chamber. If the chamber has the capability to create relative humidity conditions, an additional reference hygrometer is used to measure these conditions.
Calibration Service
Range
CMC
ClimaticChambers - Temperature Measurements
-40 °C to 140 °C
0.2 °C
ClimaticChambers - Humidity Measurements
10 %rh to 90 %rh
At 10 °C to 50 °C
0.6 %rh to 1.5 %rh
Humidity
Relative Humidity
The Humidity Laboratory uses a two-pressure generator to create reference relative humidity and air temperature conditions.
Calibration Services provided are:
Relative Humidity Sensors
Calibration of relative humidity sensors and data loggers in the range of 10 %rh to 95 % rh and air temperature between 10 °C and 70 °C.
Calibration Service
Range
CMC
Relative Humidity Sensors
10 %rh to 95 %rh
At 10 °C to 45 °C
0.4 %rh to 1.1 %rh
10 %rh to 95 %rh
At 45 °C to 70 °C
0.5 %rh to 1.7 %rh
Dew Point Meters
Calibration of dew point meters in the range of 10 %rh to 95 % rh and air temperature between 10 °C and 70 °C.
Calibration Service
Range
CMC
Dew Point Meters
10 %rh to 95 %rh
At 10 °C to 45 °C
0.3 %rh to 1.0 %rh
10 %rh to 95 %rh
At 45 °C to 70 °C
0.4 %rh to 1.6 %rh
Air Temperature Sensors
Calibration of air temperature sensors and data loggers in the range of 10°C to 70°C, in an air chamber.
Calibration Service
Range
CMC
Air Temperature Sensors
10 °C to 70 °C
0.10 °C to 0.30 °C
X
Cookies help us improve your website experience.By using our website, you agree to our use of cookies.
Confirm
Temperature
The Temperature Laboratory realises the International Temperature Scale of 1990 (ITS-90) through a set of fixed point cells (TPW, Hg, Ga, In, Sn, Zn, Al, Ag and Cu) and a set of standard platinum resistance thermometers (SPRTs).
Calibration services provided are:
Standard Platinum Resistance Thermometers
Calibration of Standard Platinum Resistance Thermometers directly at the fixed points of TPW, Hg, Ga, In, Sn, Zn and Al, covering the temperature range from -38.8344 °C to 660.323 °C.
Calibration Service
Range
Calibration and Measurement Capability (CMC)
SPRT Calibration at Fixed Points
-38.8344 °C
0.6 mK
0.01 °C
0.3 mK
29.7646 °C
0.6 mK
156.5985 °C
1.7 mK
231.928 °C
1.6 mK
419.527 °C
1.9 mK
660.323 °C
16 mK
SPRT Calibration at Fixed Points Sub Range
-38.8344 °C to 0.01 °C
0.7mK
0.01 °C to 29.7646 °C
0.7 mK
29.7646 °C to 156.5985 °C
1.8 mK
156.5985 °C to 231.928 °C
2 mK
231.928 °C to 419.527 °C
3 mK
419.527 °C to 660.323 °C
3mK to 16 mK
Platinum Resistance Thermometers
Calibration of Platinum Resistance Thermometers in comparison with SPRTs, using liquid baths to create uniform and stable temperature conditions, covering the range from -80 °C to 250 °C.
Calibration Service
Range
CMC
Platinum Resistance Thermometers
-80 °C to 80 °C
0.01
> 80 °C to 150 °C
0.015
> 150 °C to 250 °C
0.02
419.527 °C
0.005
Temperature Sensors with Display Unit/
Temperature sensors with display units are calibrated in liquid baths and in block calibrators in comparison with reference thermometers.
Calibration Service
Range
CMC
Resistance Thermometers With Display Unit
-80 °C to 80 °C
0.015 °C
> 80 °C to 150 °C
0.020 °C
> 150 °C to 250 °C
0.025 °C
> 250 °C to 400 °C
0.6 °C to 0.8 °C
Thermocouples With Display Unit
-80 °C to 150 °C
0.2 °C
> 150 °C to 250 °C
0.4 °C
> 250 °C to 400 °C
0.6 °C to 0.8 °C
Liquid In Glass Thermometers
Liquid in glass thermometers are calibrated in liquid baths from -60 °C to 250 °C by comparison with reference thermometers.
Calibration Service
Range
CMC
Liquid In Glass Thermometers
-60 °C to 250 °C
0.02 °C to 0.05 °C
Air Temperature Sensors
Calibration of air temperature sensors and data loggers in the range of 10°C to 70°C, in an air chamber.
Calibration Service
Range
CMC
Air Temperature Sensors
10 °C to 70 °C
0.10 °C to 0.30 °C
Infrared thermometers
IR Thermometers are calibrated by comparison using black body sources in the range from -150 °C to 500 °C.
Calibration Service
Range
CMC
Infrared thermometers
-15 °C to 500 °C
0.6 °C to 2.6 °C
Temperature Block Calibrators
Temperature block calibrators are calibrated using platinum resistance thermometers in the range from -30 °C to 400 °C.
Calibration Service
Range
CMC
Temperature Block Calibrators
-30 °C to 400 °C
0.2 °C
Liquid Baths
Calibration of liquid baths is carried out from -80 °C and 250 °C. The temperature profile of the bath can also be measured.
Calibration Service
Range
CMC
Liquid Baths
-80 °C to 250 °C
0.02 °C to 0.03
°C
Climatic Chambers
Climatic chambers are calibrated in the range from -40 °C to 140 °C. The temperature profile of the chamber can also be measured using 9 PRT sensors distributed in the working volume of the chamber. If the chamber has the capability to create relative humidity conditions, an additional reference hygrometer is used to measure these conditions.
Calibration Service
Range
CMC
ClimaticChambers - Temperature Measurements
-40 °C to 140 °C
0.2 °C
ClimaticChambers - Humidity Measurements
10 %rh to 90 %rh
At 10 °C to 50 °C
0.6 %rh to 1.5 %rh
Humidity
Relative Humidity
The Humidity Laboratory uses a two-pressure generator to create reference relative humidity and air temperature conditions.
Calibration Services provided are:
Relative Humidity Sensors
Calibration of relative humidity sensors and data loggers in the range of 10 %rh to 95 % rh and air temperature between 10 °C and 70 °C.
Calibration Service
Range
CMC
Relative Humidity Sensors
10 %rh to 95 %rh
At 10 °C to 45 °C
0.4 %rh to 1.1 %rh
10 %rh to 95 %rh
At 45 °C to 70 °C
0.5 %rh to 1.7 %rh
Dew Point Meters
Calibration of dew point meters in the range of 10 %rh to 95 % rh and air temperature between 10 °C and 70 °C.
Calibration Service
Range
CMC
Dew Point Meters
10 %rh to 95 %rh
At 10 °C to 45 °C
0.3 %rh to 1.0 %rh
10 %rh to 95 %rh
At 45 °C to 70 °C
0.4 %rh to 1.6 %rh
Air Temperature Sensors
Calibration of air temperature sensors and data loggers in the range of 10°C to 70°C, in an air chamber.
Calibration Service
Range
CMC
Air Temperature Sensors
10 °C to 70 °C
0.10 °C to 0.30 °C
X